Instruction Manual August 2008100/200multiranger

Pagina 2 - Weighing Systems

viiimmmmmTable of ContentsObstructions in the Sound Beam ... 2

Pagina 3 - Table of Contents

Page 90 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmCommunicationsCommunication SystemsThe MultiRanger is capable of communicating with

Pagina 4

7ML19981FB06 MultiRanger 100/200 - INSTRUCTION MANUAL Page 91mmmmmCommunicationsSmartLinxOther protocols are available through optional SmartLinx comm

Pagina 5

Page 92 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmComms InstallationCommunications Installation Wiring Guidelines• the RJ-11 cable max

Pagina 6

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 93mmmmmComms InstallationPort 1: RS-232 RJ-11 Jack To connect the unit to a PC using an RS-

Pagina 7

Page 94 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmComms InstallationTo connect the unit to a modem using an RS-485 port:Configuring Co

Pagina 8

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 95mmmmmComms InstallationFor devices connected with the Siemens Milltronics protocol, this

Pagina 9

Page 96 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmComms InstallationP778 Modem AvailableSets the MultiRanger to use an external modem.

Pagina 10

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 97mmmmmModbus Register TypesModbus Register MapFeatures affecting volume, mA input, and Ave

Pagina 11 - The MultiRanger 100 and 200

Page 98 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmModbus Register TypesThe MultiRanger was designed to make it easy for master devices

Pagina 12 - Configuration Examples

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 99mmmmmModbus Register TypesProduct ID (R40,064)This value identifies the Siemens device ty

Pagina 13 - Specifications

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 1mmmmmMultiRanger 100/200The MultiRanger 100 and 200The MultiRanger is available in two mod

Pagina 14

Page 100 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmModbus Register TypesDiscrete Inputs (R41,070) This table shows the current status

Pagina 15

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 101mmmmmModbus Register TypesPump OFF Setpoint (R41,430 – R41,435) The OFF setpoint level (

Pagina 16 - Panel Mount

Page 102 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmModbus Register TypesParameter Access (R43,998 – R46,999)Parameter values are given

Pagina 17

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 103mmmmmModbus Register TypesReading ParametersTo read parameter values, follow the steps l

Pagina 18 - Installation

Page 104 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmModbus Register Types•from Transducer OneSend the integer value 01008 to register 4

Pagina 19 - Mounting Instructions

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 105mmmmmModbus Register TypesFormat Words (R46,000 to R46,999)Format words are unsigned int

Pagina 20 - 1, Div 2 Applications

Page 106 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmModbus Register TypesThe decimal offset indicates how the remote system must interp

Pagina 21

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 107mmmmmData TypesData TypesThe MultiRanger parameters do not always use integers to hold v

Pagina 22 - Mounting the Enclosure

Page 108 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmData TypesFor example, if R41,442 is read as a UINT32, the 32 bits would look like

Pagina 23 - Installing the Battery

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 109mmmmmData TypesText MessagesIf a device parameter returns a text message, that message i

Pagina 24 - Installing SmartLinx Card

Page 2 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmMultiRanger 100/200If you have any questions, comments, or suggestions about the manu

Pagina 25 - Please note:

Page 110 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmData TypesRelay Function Codes (P111 Only) Please note that the MultiRanger 200 off

Pagina 26 - Terminal Board

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 111mmmmmError HandlingError HandlingModbus ResponsesWhen polled by a Modbus Master, a slave

Pagina 27 - Transducers

Page 112 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmError Handling• If the host attempts to write one or more registers that are out of

Pagina 28 - Temperature Sensor

7ML19981FB06 MultiRanger 100/200 - INSTRUCTION MANUAL Page 113mmmmmCommunicationsCommunication TroubleshootingGenerally1. Check the following:• There

Pagina 29 - MultiRanger

Page 114 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmCommunication Appendix ACommunication Appendix A: Single Parameter Access (SPA)This

Pagina 30 - Digital Communications

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 115mmmmmCommunication Appendix AWriting ParametersTo set parameters through Modbus do the f

Pagina 31 - Discrete Inputs

Page 116 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmCommunication Appendix AError CodesThe error codes returned in the format area are

Pagina 32 - Operating the MultiRanger

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 117mmmmmParametersParameter ReferenceThe MultiRanger is configured through its parameters,

Pagina 33

Page 118 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersIndex typesName Description # of indexesGlobal This parameter applies to

Pagina 34 - Status Parameters

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 119mmmmmParametersP000 LockSecures the MultiRanger from changes via the handheld programmer

Pagina 35 - Controlling the Display

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 3mmmmmSpecificationsSpecificationsPowerAC version• 100-230 V AC ± 15%, 50 / 60 Hz, 36 VA (1

Pagina 36 - Parameter Indexing

Page 120 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersFor DPD and DPA Programming [MR 200]Please note that this feature is only

Pagina 37

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 121mmmmmParametersP003 Maximum Process SpeedDetermines level change reaction.Use a setting

Pagina 38 - Programmer Keys

Page 122 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP005 UnitsSpecifies measurement units used for dimensional values. Changi

Pagina 39

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 123mmmmmParametersP007 SpanSets the range levels to be measured. Span is preset for a value

Pagina 40 - Dolphin Plus Toolbar Buttons

Page 124 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersPrimary Index LevelValues# Shape Description0*volume calculation not requ

Pagina 41 - Device Description

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 125mmmmmParametersP051 Maximum Volume [MR 200]For Readings in volumetric units (rather than

Pagina 42 - Activating the MultiRanger

Page 126 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP052 Tank Dimension A [MR 200]Dimension A as used in P050 Tank Shape.Ente

Pagina 43 - Parameters P925–P927

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 127mmmmmParametersP055 Volume Breakpoints and Characterization (Universal Volume Calculatio

Pagina 44 - Display Readout

Page 128 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersIn RUN mode, the decimal position adjusts to prevent the number of digits

Pagina 45 - Operation

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 129mmmmmParametersBackup Level OverrideUse this feature to override the ultrasonic reading

Pagina 46

Page 4 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmSpecificationsRange• 0.3 m (1 ft) to 15 m (50 ft), dependent on transducerAccuracy• 0

Pagina 47 - Primary and Secondary Indexes

Page 130 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersPlease note the following:• enter value in current units (as selected in

Pagina 48 - • P054 [2,1] = 8m

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 131mmmmmParametersFailsafe (P070 to P072) P070 Failsafe TimerThe time for invalid measureme

Pagina 49 - Dual Point Models

Page 132 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersSelecting HI, LO, or HOLd1. Press FUNCTION to display the Auxiliary Fun

Pagina 50 - Troubleshooting Appendix

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 133mmmmmParametersControl Functions (P111)Each relay can be configured independently to tak

Pagina 51 - Relay Function

Page 134 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP110 Level SourceSets the level source on which the indexed relay is assi

Pagina 52 - 2 % of span

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 135mmmmmParametersMultiRanger 100MultiRanger 200Values For P111Control Type#11.When reading

Pagina 53 - Relay Related Parameters

Page 136 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP112 Relay ON SetpointSets the process point at which the relay changes f

Pagina 54 - Relay Activation

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 137mmmmmParametersP113 Relay OFF SetpointSets the process point at which the relay returns

Pagina 55

Page 138 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersFor IN-BOUNDS and OUT-OF-BOUNDS Relay Functions (P111 = 2 and 3 respectiv

Pagina 56 - Preset Applications

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 139mmmmmParametersPositive LogicIn software, all relays are programmed the same way, with O

Pagina 57

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 5mmmmmSpecificationsFixed• 0.17 % per °C deviation from programmed valueOutputsTransducer d

Pagina 58

Page 140 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersThe delay between pump starts is set by P132 Pump Start Delay.This only a

Pagina 59

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 141mmmmmParametersRelay Failsafe is only available for the following relay functions (P111)

Pagina 60 - Verifying the mA Range

Page 142 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP131 Pump Run-On Duration [MR 200]Sets the number of seconds that the pum

Pagina 61 - Volume [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 143mmmmmParametersP136 Wall Cling Reduction [MR 200]Varies the upper and lower setpoints to

Pagina 62 - Example Chart

Page 144 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersFlush Systems (P170 to P173) [MR 200]Please note that these parameters ap

Pagina 63 - MultiRanger 200 only

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 145mmmmmParametersP171 Flush Cycles [MR 200]Sets the number of pump cycles requiring flush

Pagina 64 - Value Description

Page 146 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersmA Output (P200 to P219)P200 mA Output Range Determines the mA output ran

Pagina 65 - Rate [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 147mmmmmParametersMultiRanger 200P202 mA Output Allocation Sets the input source from which

Pagina 66 - Temperature [MR 200]

Page 148 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersEnter the Point Number the mA output is based on. This value depends on w

Pagina 67 - Loss of Echo (LOE)

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 149mmmmmParametersIndependent mA Setpoints (P210 and P211) Use these features to reference

Pagina 68

Page 6 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmSpecificationsInputsmA (analog) (1) [MR 200 only]• 0-20 or 4-20 mA, from alternate de

Pagina 69

Page 150 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP212 mA Output Min LimitSets the minimum mA output value (in mA) to be pr

Pagina 70

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 151mmmmmParametersP215 20 mA Output Trim Calibrates the 20 mA output.Adjust this value so t

Pagina 71 - Other Pump Control Algorithms

Page 152 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP250 mA Input Range [MR 200]Shows the mA input range of the connected mA

Pagina 72 - [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 153mmmmmParametersP253 Input Filter Time Constant [MR 200]Shows the time constant used in t

Pagina 73 - Optional Pump Controls

Page 154 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP270 Discrete Input FunctionSets how discrete signals are interpreted by

Pagina 74

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 155mmmmmParametersP300 Temperature, Transducer MaximumShows the highest temperature encount

Pagina 75 - Set in RUN Mode

Page 156 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP303 Temperature, Sensor MinimumShows the lowest temperature encountered,

Pagina 76 - Reducing Wall Cling [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 157mmmmmParametersP309 Pump RUN TimeDisplays the amount of time in minutes since a relay wa

Pagina 77 - Grouping Pumps [MR 200]

Page 158 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersFlow Records (P320 and P321) [MR 200]Please note that these parameters ap

Pagina 78 - Tracking Pump Usage

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 159mmmmmParametersTotalizer units are dependent upon programming. Enter zero 0 (if required

Pagina 79 - Setting a Rake Control

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 7mmmmmSpecificationsCable• Do not use coaxial cable for transducer (see General Appendix F:

Pagina 80 - Set Relay 1 (Operate Rake)

Page 160 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP330 Profile RecordRecords profiles for later viewing.In addition to bein

Pagina 81

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 161mmmmmParametersP331 Auto Record EnableUse to enable/disable the Auto Profile Record func

Pagina 82 - Flow Sampler

Page 162 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersAuto Record ON and OFF Setpoints (P334 to P337)Use Auto Record ON Setpoin

Pagina 83

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 163mmmmmParametersP336 Auto Record Filling / EmptyingUse this feature to restrict Auto Prof

Pagina 84 - Setting Zero Head

Page 164 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersInstallation Records (P340 to P342) P340 Date of ManufactureView the date

Pagina 85 - Setting Totalized Volume

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 165mmmmmParametersOpen Channel Monitoring (P600 to P621) [MR 200]Open Channel Monitoring (O

Pagina 86 - Parameter Index Value

Page 166 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersThe MultiRanger is pre-programmed for common PMD flow calculations. If yo

Pagina 87 - Palmer Bowlus Flume

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 167mmmmmParametersExample ExponentsP602 Primary Measuring Device Dimensions [MR 200]The dim

Pagina 88

Page 168 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP603 Maximum Head [MR 200]The level value associated with Maximum Flow, i

Pagina 89 - Applicable Weir Profiles

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 169mmmmmParametersExampleP605 Zero Head [MR 200]The distance above Empty (P006) in Units (P

Pagina 90

Page 8 MultiRanger 100/200– INSTRUCTION MANUAL 7ML19981FB06mmmmmInstallationInstallationMountingMounting LocationsRecommended• Ambient temperature is

Pagina 91 - Leopold Lagco Flume

Page 170 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP606 Time Units [MR 200]Determines the units used to display current flow

Pagina 92 - Cut Throat Flume

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 171mmmmmParametersP608 Flowrate Units [MR 200]The volume units used to display total flow.T

Pagina 93

Page 172 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP610 Head Breakpoints [MR 200]The head breakpoints for which flowrate is

Pagina 94 - Example Weirs

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 173mmmmmParametersP620 Low Flow Cutoff [MR 200]Eliminates totalizer activity for flows at o

Pagina 95 - Testing the Configuration

Page 174 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersPumped Volume Totalizer (P622) [MR 200]The Pumped Volume Totalizer is a M

Pagina 96

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 175mmmmmParametersTotalizer (P630 to P645) [MR 200]The Totalizer features apply to MultiRan

Pagina 97

Page 176 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP633 LCD Totalized Decimal Position [MR 200]Enter the maximum number of d

Pagina 98 - Troubleshooting

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 177mmmmmParametersEnter the factor (powers of 10 only) by which actual volume is divided, p

Pagina 99

Page 178 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersThis parameter is relevant only if Operation is set to OCM (P001 = 6).Ent

Pagina 100 - Communication Ports

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 179mmmmmParametersP650 Offset CalibrationCalibrates Empty (P006) if the reported level is c

Pagina 101 - Dolphin Plus

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 9mmmmmInstallationMounting InstructionsThe wall mount and panel mount units install differe

Pagina 102 - Communications Installation

Page 180 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersUsing Sound Velocity CalibrationEnsure a steady level at some low value (

Pagina 103 - Port 2: RS-485

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 181mmmmmParametersP654 Velocity at 20°CThis value is used to automatically calculate Sound

Pagina 104 - P771 Network Address

Page 182 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersIn gasses other than air, the temperature variation may not correspond wi

Pagina 105 - The serial port parity

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 183mmmmmParametersP664 TemperatureView the transducer temperature in °C.Value is displayed

Pagina 106 - Parameter Access

Page 184 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP701 Max Empty RateAdjusts the MultiRanger response to decreases in the a

Pagina 107 - Appendix A

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 185mmmmmParametersP703 Emptying IndicatorThe empty rate required to activate the LCD Emptyi

Pagina 108 - Location

Page 186 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP705 Rate Update TimeThe time period (in seconds) over which the material

Pagina 109 - Parameter Reference

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 187mmmmmParametersP708 Volume Rate Display [MR 200]The rate of change of volume in percent

Pagina 110 - Relay Data Address

Page 188 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP711 Echo LockUse this feature to select the measurement verification pro

Pagina 111 - Parameter Reference

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 189mmmmmParametersP712 Echo Lock SamplingThe sampling criterion sets the number of consecut

Pagina 112 - Global Index Method

© Siemens Milltronics Process Instruments Inc. 2009 Safety Guidelines: Warning notices must be observed to ensure personal safety as well as that

Pagina 113 - Reading Parameters

Page 10 MultiRanger 100/200– INSTRUCTION MANUAL 7ML19981FB06mmmmmInstallationCable routed through a conduit:1. Remove the four mounting screws holding

Pagina 114 - Writing Parameters

Page 190 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersThe Echo Lock Window is a distance window (units P005) centred on the ech

Pagina 115 - Format Registers

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 191mmmmmParametersP728 Shot DelayThe delay, in seconds, between transducer shots.Use this i

Pagina 116 - Example

Page 192 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersIf necessary, enter the Parameter Number to default in the Auxiliary Read

Pagina 117

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 193mmmmmParametersP735 BacklightControls the LCD backlighting.The backlight can be forced o

Pagina 118 - Word Order (R40,062)

Page 194 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersSmartLinx Reserved (P750 to P769) These parameters are reserved for optio

Pagina 119 - Text Messages

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 195mmmmmParametersP772 Baud RateThe communication rate with the master device.This specifie

Pagina 120

Page 196 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP775 Stop BitsThe number of bits between the data bits.P778 Modem Availab

Pagina 121 - Error Handling

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 197mmmmmParametersP782 Parameter Index LocationDetermines where index information is stored

Pagina 122

Page 198 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP791 Hardware Error CodeIndicates the precise cause of Fail or ERR1 condi

Pagina 123 - Communication Troubleshooting

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 199mmmmmParametersP799 Communications ControlEnables the read/write access to parameters vi

Pagina 124 - Parameter Access (SPA)

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 11mmmmmInstallationPanel MountInstalling the panel mount unit requires making a cutout in t

Pagina 125 - Format Register

Page 200 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP801 Range ExtensionAllows the material level to fall below the Empty set

Pagina 126 - Error Codes

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 201mmmmmParametersP803 Shot / Pulse ModeDetermines what type of ultrasonic shots are fired.

Pagina 127

Page 202 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP805 Echo ConfidenceDisplays the echo confidence of the measurement echo

Pagina 128 - Index types

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 203mmmmmParametersAdvanced Echo Processing (P815 to P825) The following parameters are for

Pagina 129 - Quick Start (P001 to P007)

Page 204 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP820 AlgorithmChooses the algorithm to generate the measured value from t

Pagina 130 - Specifies material type

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 205mmmmmParametersP821 Spike FilterDampens spikes in the echo profile to reduce false readi

Pagina 131 - P004 Transducer

Page 206 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP823 Reform Echo Smoothes jagged peaks in the echo profile.Use this featu

Pagina 132 - P006 Empty

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 207mmmmmParametersAdvanced TVT Adjustment (P830 to P835) The following parameters are for a

Pagina 133 - Maximum Volume (P051)

Page 208 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP832 TVT Shaper AdjustAllows manual adjustment of the TVT curve. Use this

Pagina 134 - # Shape Description

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 209mmmmmParametersP834 TVT Start DurationUse this feature in conjunction with TVT Start Min

Pagina 135 - P051 Maximum Volume [MR 200]

Page 12 MultiRanger 100/200– INSTRUCTION MANUAL 7ML19981FB06mmmmmInstallationPanel Mount DimensionsMounting the EnclosureOnce cutout is complete and m

Pagina 136 - [MR 200]

Page 210 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP837 Auto False-Echo SuppressionUse P837 and P838 together, to set MultiR

Pagina 137 - Volume Characterization

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 211mmmmmParametersP838 Auto False-Echo Suppression DistanceDefines the range of Auto False-

Pagina 138 - P062 Offset Reading

Page 212 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersAdvanced Shot Adjustment (P840 to P852) These parameters are for Siemens

Pagina 139 - Backup Level Override

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 213mmmmmParametersP842 Short Shot FrequencyAdjust the short shot transmit pulse frequency (

Pagina 140 - P069 Password

Page 214 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP845 Long Shot WidthAdjust the width (in ms) of the long shot transmit pu

Pagina 141 - Failsafe

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 215mmmmmParametersP852 Short Shot RangeEnter the maximum distance in Units (P005) to be mea

Pagina 142 - Relays (P100 to P119)

Page 216 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP902 WatchdogPress ENTER to put the CPU into an infinite loop to test t

Pagina 143 - Setpoints (P112, P113)

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 217mmmmmParametersP908 ScannerPress ENTER to cycle the scanner relay while firing the tra

Pagina 144 - P111 Relay Control Function

Page 218 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP913 Sensor TemperatureAccess this parameter to display the temperature i

Pagina 145 - MultiRanger Communications

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 219mmmmmParametersReading Measurements by OperationP921 Material MeasurementThe distance in

Pagina 146

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 13mmmmmInstallationMultiRanger BoardInstalling the BatteryThe battery (Rayovac BR2032) has

Pagina 147 - P116 Dead Band [MR 200]

Page 220 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersP924 Volume Measurement [MR 200]The calculated vessel capacity in Max Vol

Pagina 148 - Relay Control Function

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 221mmmmmParametersP927 Distance MeasurementThe distance between the surface and the transdu

Pagina 149 - Pump Control

Page 222 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParametersNotes:

Pagina 150 - P129 Relay Failsafe

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 223mmmmmAppendix A: Index TypesAppendix A: Index TypesIndex typesName Description # of inde

Pagina 151

Page 224 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix B: Tech ReferenceAppendix B: Technical ReferenceTransmit PulseThe transmit

Pagina 152

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 225mmmmmAppendix B: Tech ReferenceTVT (Time Varying Threshold) curvesA TVT curve describes

Pagina 153 - P137 Pump Group [MR 200]

Page 226 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix B: Tech ReferenceDistance CalculationTo calculate the transducer to materi

Pagina 154 - P170 Flush Pump [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 227mmmmmAppendix B: Tech ReferenceScanningMultiRanger 100/200When echo processing is comple

Pagina 155 - P173 Flush Duration [MR 200]

Page 228 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix B: Tech ReferenceUniversal, Curved [MR 200]Please note that the Universal,

Pagina 156 - P201 mA Output Function

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 229mmmmmAppendix B: Tech ReferenceUniversal, Linear [MR 200]Please note that the Universal,

Pagina 157 - P202 mA Output Allocation

Page 14 MultiRanger 100/200– INSTRUCTION MANUAL 7ML19981FB06mmmmmInstallationInstalling SmartLinx Card SmartLinx cards are generally pre-installed. If

Pagina 158

Page 230 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix B: Tech ReferenceFor combination curves, enter at least 2 breakpoints imme

Pagina 159 - P211 20 mA Output Setpoint

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 231mmmmmAppendix C: TroubleshootingAppendix C: TroubleshootingIf the setup becomes too conf

Pagina 160 - Calibrates the 4 mA output

Page 232 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix C: TroubleshootingNoise ProblemsIncorrect readings can be the result of no

Pagina 161 - Calibrates the 20 mA output

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 233mmmmmAppendix C: TroubleshootingNon-Transducer Noise SourcesRemove all input and output

Pagina 162 - P250 mA Input Range [MR 200]

Page 234 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix C: TroubleshootingMeasurement DifficultiesIf the Failsafe Timer (P070) exp

Pagina 163

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 235mmmmmAppendix C: TroubleshootingTo display Echo Confidence in the RUN modePress and ho

Pagina 164 - P270 Discrete Input Function

Page 236 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix C: Troubleshootinglarger diameter or shorter length nozzle, bevel the insi

Pagina 165

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 237mmmmmAppendix C: TroubleshootingIf multiple echoes appear on the Echo Profile, typical o

Pagina 166 - Sensor (if applicable)

Page 238 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix C: TroubleshootingRing down times that extend past the near blanking area

Pagina 167 - P312 Pump Run Ons [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 239mmmmmAppendix D: Pump ControlAppendix D: Pump Control ReferencePlease note that some pum

Pagina 168 - P321 Flow Min [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 15mmmmmWiringWiringPlease note:• Verify that all system components are installed in accorda

Pagina 169 - P323 LCD Total High [MR 200]

Page 240 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix D: Pump ControlAll of these algorithms can be used to start multiple pumps

Pagina 170 - P330 Profile Record

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 241mmmmmAppendix D: Pump ControlThe following table shows relay status when each setpoint i

Pagina 171 - P331 Auto Record Enable

Page 242 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix D: Pump ControlRelay TableService Ratio Duty Assist (P111 = 54) [MR 200]Se

Pagina 172 - P335 Auto Record OFF Setpoint

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 243mmmmmAppendix D: Pump ControlService Ratio Duty Backup (P111 = 55) [MR 200]Selects the l

Pagina 173 - P337 Auto Record LOE Time

Page 244 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix D: Pump ControlFlush Device (P170 to P173)Operates a flush valve or specia

Pagina 174 - P342 Start Ups

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 245Appendix E: Updating SoftwareAppendix E: Updating SoftwareUpdating SoftwareTo update the

Pagina 175 - Open Channel Monitoring

Page 246 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix F: Upgrading Appendix F: UpgradingThe following procedure will assist you

Pagina 176 - P601 Flow Exponent [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 247mmmmmAppendix F: Upgrading Connecting a transducer with RG62 coaxial extension cableIf y

Pagina 177 - Example Exponents

Page 248 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix F: Upgrading MultiRanger Plus to MultiRanger 100/200 ParametersThe MultiRa

Pagina 178 - P604 Maximum Flow [MR 200]

7ML19981FBO6 MultiRanger 100/200 – INSTRUCTION MANUAL Page 249Appendix GAppendix G: Conduit Entry for Class 1, Div 2 ApplicationsProduct Group23650314

Pagina 179 - P605 Zero Head [MR 200]

Page 16 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmWiringTerminal BoardThe terminal board on the MultiRanger allows all inputs and outp

Pagina 180 - P606 Time Units [MR 200]

Page 250 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FBO6Appendix GProduct Group2365031400S. NGUYENPETERBOROUGH14/ AUG / 02ULTRASONICSFOR CONSTRU

Pagina 181 - P608 Flowrate Units [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 251mmmmmAppendix H: Software Rev.Appendix H: Software Revision HistoryMultiRanger 100Soft-w

Pagina 182 - Universal Calculation Support

Page 252 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix H: Software Rev.1.041.05Feb. 11, 2003Feb. 23, 2003• Product enhancements t

Pagina 183 - P621 Auto Zero Head [MR 200]

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 253mmmmmAppendix H: Software Rev.1.08 Nov. 11, 2004• Improvements made as a result of Sieme

Pagina 184

Page 254 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix H: Software Rev.MultiRanger 200Soft-ware Rev.Date Changes1.00 Nov. 09, 200

Pagina 185 - Example:

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 255mmmmmAppendix H: Software Rev.1.041.05Feb. 11, 2003Feb. 21, 2003• Product enhancements t

Pagina 186

Page 256 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAppendix H: Software Rev.1.08 Aug. 30, 2004• Improvements made as a result of Sieme

Pagina 187

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 257mmmmmParameter ChartsProgramming Charts Parameter Altered Values for Indices/Point Numbe

Pagina 188 - P645 Relay Duration [MR 200]

Page 258 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParameter ChartsP113 Relay OFF SetpointP115 Relay Interval Setpoint [MR 200]P116 De

Pagina 189 - P650 Offset Calibration

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 259mmmmmParameter ChartsP219 mA Output Failsafe[MR 200]mA InputP250 mA Input Range [MR 200]

Pagina 190 - P653 Velocity

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 17mmmmmWiringTransducersA 0.1 µF (100V or greater) capacitor is included with the MultiRang

Pagina 191 - P660 Temp Source

Page 260 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParameter ChartsP334 Auto Record ON SetpointP335 Auto Record OFF SetpointP336 Auto

Pagina 192 - P661 Temp Fixed

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 261mmmmmParameter ChartsRange CalibrationP650 Offset CalibrationP651 Sound Velocity Calibra

Pagina 193 - Readings in Run Mode

Page 262 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParameter ChartsP737 Primary Reading [MR 200]P741 Communications TimeoutSmartLinx R

Pagina 194 - P702 Filling Indicator

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 263mmmmmParameter ChartsP824 (reserved)P825 Echo Marker TriggerAdvanced TVT AdjustmentP830

Pagina 195 - P704 Rate Filter

Page 264 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmParameter ChartsP920 Reading MeasurementP921 Material MeasurementP922 Space Measure

Pagina 196 - Readings in RUN Mode

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 265mmmmmIndexIndexAaccuracy ...4

Pagina 197 - P710 Fuzz Filter

Page 266 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmIndexLeopold Lagco ... 81Palmer Bowlus ...

Pagina 198 - P711 Echo Lock

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 267mmmmmIndexoverview ... 102reading ...

Pagina 199 - P713 Echo Lock Window

Page 268 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmIndexregister map ... 97discrete i

Pagina 200 - P727 Scan Delay

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 269mmmmmIndexcharacterization chart ... 52dimensions ...

Pagina 201 - Hand Programmer

Page 18 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmWiringTemperature SensorAccurate temperature readings are critical to accurate level

Pagina 202

*7ml19981FB06*Rev. 6.0www.siemens.com/processautomationSiemens Milltronics Process Instruments Inc.1954Technology Drive, P.O. Box 4225Peterborough, ON

Pagina 203 - P741 Communications Timeout

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 19mmmmmWiringLevel System SynchronizationWhen using multiple ultrasonic level monitors, be

Pagina 204 - Communications

immmmmTable of ContentsTable of ContentsThe MultiRanger 100 and 200 ...

Pagina 205

Page 20 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmWiringDigital CommunicationsWiring the MultiRanger for communications allows it to b

Pagina 206 - P779 Modem Inactivity Timeout

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 21mmmmmWiringDiscrete InputsDiscrete inputs have a positive and negative terminal. Requires

Pagina 207 - P790 Hardware Error

Page 22 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationOperating the MultiRangerThe MultiRanger has two modes of operation: RUN an

Pagina 208 - P795 SmartLinx Protocol

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 23mmmmmOperationIcons indicating index type (Item 1) edited in PROGRAM mode:Readings in RUN

Pagina 209 - P800 Near Blanking

Page 24 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationStatus ParametersStatus parameters give the operating status of the MultiRa

Pagina 210

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 25mmmmmOperationControlling the DisplayRUN mode provides numerous parameters and variables

Pagina 211 - P804 Confidence Threshold

Page 26 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationSetting the Default Auxiliary ReadingTo maintain a constant variable displa

Pagina 212 - P807 Noise

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 27mmmmmOperationPROGRAM ModeThe MultiRanger is programmed by setting its parameters to matc

Pagina 213 - P816 Echo Time Raw

Page 28 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationProgrammer KeysKeys Programming Mode Run Mode1 8-digit Totalizer (toggle) [

Pagina 214 - P820 Algorithm

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 29mmmmmOperationDolphin Plus(compatible with product software versions 1.06 and earlier)1Us

Pagina 215 - P822 Narrow Echo Filter

iimmmmmTable of ContentsHand Programmer ...

Pagina 216 - P825 Echo Marker Trigger

Page 30 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationDolphin Plus Toolbar ButtonsThe toolbar buttons provide quick access to Dol

Pagina 217 - Selects the TVT Curve used

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 31mmmmmOperationSIMATIC Process Device Manager (PDM)(compatible with product software versi

Pagina 218 - Dolphin Plus PC Software

Page 32 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationActivating the MultiRangerAll the instructions in the following procedures

Pagina 219 - P835 TVT Slope Min

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 33mmmmmOperationSecurityThe Lock parameter P000 secures the MultiRanger against parameter c

Pagina 220 - (or when P837 = 0)

Page 34 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationDefault ValuesParameter default values are indicated with an * in the param

Pagina 221 - P839 TVT Hover Level

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 35mmmmmOperationChanging Parameters(Dolphin Plus)The other method for changing parameter va

Pagina 222 - P841 Long Shot Number

Page 36 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationParameter IndexingParameters are indexed when they apply to more than one i

Pagina 223 - P844 Short Shot Width

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 37mmmmmOperationPrimary and Secondary IndexesPrimary Index: relates to direct input or outp

Pagina 224 - P851 Short Shot Floor

Page 38 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationExample [MR 200 only]P054 provides up to 32 breakpoint levels used with P05

Pagina 225 - View the EPROM Rev. #

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 39mmmmmOperation3Average or Differential [MR 200 only]For differential or average operation

Pagina 226 - P906 RS-232 Port

iiimmmmmTable of ContentsmA I/O ...

Pagina 227 - P912 Transducer Temperature

Page 40 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOperationAverage or Differential [MR 200 only]For differential or average operation

Pagina 228 - Configuration

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 41mmmmmRelaysRelaysRelays are the primary controls of external devices such as pumps or ala

Pagina 229 - P923 Distance Measurement

Page 42 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmRelaysTemperature [MR 200]In high alarm, the function goes on when the temperature r

Pagina 230

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 43mmmmmRelaysRelay Status – Non Run Modes When the fail-safe timer expires, pump control re

Pagina 231 - Master Reset (P999)

Page 44 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmRelaysP111–Control FunctionsThe control function energizes the relay coils. When the

Pagina 232

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 45mmmmmRelays[MR 100]: Functions affected by setpoint are configured by parameters that det

Pagina 233 - Appendix A: Index Types

Page 46 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmRelaysPreset ApplicationsPreset applications set up the relay parameters to predeter

Pagina 234 - Echo Processing

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 47mmmmmBackup Level OverrideBackup Level OverrideBackup level override provides the option

Pagina 235 - Auto False-Echo Suppression

Page 48 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmDiscrete InputsDiscrete InputsWiring the Discrete InputsNormal state is standard ope

Pagina 236 - Sound Velocity

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 49mmmmmmA I/OmA I/OTo integrate the MultiRanger with other equipment, use the mA input and

Pagina 237 - Volume Calculation [MR 200]

ivmmmmmTable of ContentsSetting the Common Parameters ...70

Pagina 238 - Flow Calculation

Page 50 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmmA LI/OCalibrating 4 mA Output1. Connect the mA receiving device to the MultiRanger.

Pagina 239 - Typical Flow Characterization

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 51mmmmmVolumeVolume [MR 200]Volume is a feature of the MultiRanger 200 only. Volume is used

Pagina 240 - Maximum Process Speed

Page 52 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmVolumeCharacterization Chart [MR 200]If you cannot use a pre-defined tank, then use

Pagina 241 - Appendix C: Troubleshooting

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 53mmmmmVolumeMultiRanger 200 onlyParameter Transducer Index Value DescriptionP054 110.0Det

Pagina 242 - Non-Transducer Noise

Page 54 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAlarmsAlarmsLevelThe level alarm is the most common. Use this alarm to warn you when

Pagina 243 - Reducing Acoustical Noise

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 55mmmmmAlarmsSetting Simple Level AlarmsTo set Relay Five to a standard level alarm (Hi Hi,

Pagina 244 - Measurement Difficulties

Page 56 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmAlarmsIn Bounds/ Out of Bounds Range [MR 200]Use the bounded range alarms to detect

Pagina 245 - P832: TVT Shaper

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 57mmmmmAlarmsWith P112 and P113, you can set a high alarm (P112 > P113) or a low alarm (

Pagina 246 - Fixed Reading

Page 58 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmPump ControlPump ControlSetting a Pump Down GroupExample: Sewage Wet WellSetting a g

Pagina 247 - Transducer Ringing

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 59mmmmmPump ControlSet Relays to ALTERNATE DUTY ASSIST Set the ON SetpointsSet the OFF Setp

Pagina 248

vmmmmmTable of ContentsCommunications Installation ...

Pagina 249 - Pump Control Algorithms

Page 60 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmPump ControlSet the Common Parameters Prerequisite: Substitute the details of your a

Pagina 250 - Off 0 Off Off Off

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 61mmmmmPump ControlOther Pump Control AlgorithmsSet Relays to ALTERNATE DUTY BACKUP [MR 200

Pagina 251

Page 62 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmPump ControlSet Relays to FIXED DUTY BACKUP [MR 200]Set the Relay ON Setpoints [MR 2

Pagina 252

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 63mmmmmPump ControlSet Relays to FIRST IN FIRST OUT (FIFO) ASSIST [MR 200]Set the Relay ON

Pagina 253 - Other Pump Controls [MR 200]

Page 64 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmPump ControlWhen the first ON setpoint is reached, the pumps will start, one by one,

Pagina 254 - Flush Device (P170 to P173)

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 65mmmmmPump ControlTotalizing Pumped Volume [MR 200]Prerequisite: the volume of the vessel

Pagina 255 - Appendix E: Updating Software

Page 66 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmPump ControlSetting a Pump to Run On [MR 200]When you need to pump below the normal

Pagina 256

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 67mmmmmPump ControlExample [MR 200]:A range of 0.5 meters is used to vary the setpoint. The

Pagina 257 - Installations)

Page 68 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmPump ControlExample: The flush valve connects to Relay Four and the watched pump is

Pagina 258 - MultiRanger 100/200

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 69mmmmmRake (Screen) ControlMR 200Rake (Screen) Control [MR 200]This feature is only availa

Pagina 259 - Div 2 Applications

vimmmmmTable of ContentsError Handling ...

Pagina 260 - Appendix G

Page 70 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06Rake (Screen) Control MR 200Setting the Common ParametersPrerequisite: Substitute the det

Pagina 261 - MultiRanger 100

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 71mmmmmExternal Totalizers and Flow Samplers – MR 200External Totalizers and Flow Samplers

Pagina 262 - longer handled in

Page 72 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmExternal Totalizers and Flow Samplers – MR 200The source of units depends on the ope

Pagina 263 - Date Changes

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 73mmmmmOpen Channel MonitoringMR 200Open Channel Monitoring (OCM) [MR 200]An OCM installati

Pagina 264 - MultiRanger 200

Page 74 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOpen Channel MonitoringMR 200Setting Zero HeadMany PMDs start flowing higher than th

Pagina 265

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 75mmmmmOpen Channel MonitoringMR 200Setting Totalized VolumeTo display the totalized volume

Pagina 266

Page 76 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOpen Channel MonitoringMR 200BS-3680 / ISO 4359 Rectangular FlumeParameter Index Val

Pagina 267 - Programming Charts

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 77mmmmmOpen Channel MonitoringMR 200Palmer Bowlus FlumeApplication Information• Sized by pi

Pagina 268 - # Name 1 2 3

Page 78 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOpen Channel MonitoringMR 200H Flume• Sized by maximum depth of flume• Approach is p

Pagina 269

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 79mmmmmOpen Channel MonitoringMR 200PMDs with Exponential Flow to Head FunctionFor Primary

Pagina 270

viimmmmmTable of ContentsTemperature Compensation (P660 to P664) ...18

Pagina 271

Page 80 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOpen Channel MonitoringMR 200Non-Applicable Weir ProfilesFlows through these weirs c

Pagina 272

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 81mmmmmOpen Channel MonitoringMR 200Leopold Lagco FlumeApplication Information• Designed to

Pagina 273

Page 82 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOpen Channel MonitoringMR 200Cut Throat FlumeApplication Information• Similar to Par

Pagina 274

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 83mmmmmOpen Channel MonitoringMR 200Universal Calculation SupportWhen the primary measuring

Pagina 275

Page 84 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmOpen Channel MonitoringMR 200Example FlumesThese example flumes would both require a

Pagina 276

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 85mmmmmTesting the ConfigurationTesting the ConfigurationAfter programming the unit, you mu

Pagina 277

Page 86 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmTesting the ConfigurationThe effect of the ARROW key is determined by the state (rat

Pagina 278

7ML19981FB06 MultiRanger 100/200 – INSTRUCTION MANUAL Page 87mmmmmTesting the ConfigurationI/O CheckoutAfter the unit is installed, test to verify the

Pagina 279

Page 88 MultiRanger 100/200 – INSTRUCTION MANUAL 7ML19981FB06mmmmmTesting the ConfigurationMonitor system performance carefully, under all anticipated

Pagina 280 - *7ml19981FB06*

7ML19981FB06 MultiRanger 100/200 - INSTRUCTION MANUAL Page 89mmmmmCommunicationsMultiRanger CommunicationsMultiRanger Communication SystemsThe MultiRa